Concretely, purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.
Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.