How to read a certificate of analysis

Sixteen fields, in the order they matter, and what the absence of each one lets you infer. Plus the arithmetic that reconciles gross vial mass with label claim.

The one misreading that matters most. Chromatographic purity and quantified content are different measurements answering different questions. Purity is the fraction of detected material that is your target. Content is how many milligrams are in the vial. A certificate can honestly report 99.2 per cent purity on a vial containing six milligrams of a nominal ten, and nothing about the purity figure constrains that. If you read one thing on this page, read this paragraph twice.

The fields, in the order they matter

#FieldWhat it should sayWhat its absence implies
1Lot or batch numberAn identifier that also appears physically on the vial, not only on the box.You cannot tie this document to the object in your hand. This is the most consequential omission on the list.
2Date of analysisA date after the manufacturing date.If analysis precedes manufacture, the certificate belongs to a different lot. This happens more often than it should.
3Product identity and claimed contentThe compound, and the nominal quantity per vial.Without a nominal figure there is nothing for a content assay to be measured against.
4Method sectionColumn chemistry, dimensions, mobile phase, gradient, flow rate, detection wavelength, injection volume.The purity number is not reproducible and not comparable to your own independent result. Treat it as a claim.
5Purity, as area %A figure with the wavelength stated. 214 nm and 280 nm give different numbers on the same injection.A purity figure without a wavelength is incomplete by construction.
6ChromatogramAn actual trace, with the baseline and the impurity peaks visible.You cannot see whether the 1.5 per cent of non-target material is one impurity or eleven, which is the more informative question.
7Impurity tableRetention times and area percentages for each resolved impurity above the reporting threshold.You are being given a total without a distribution.
8Identity confirmationObserved mass against theoretical, with the charge state and whether it is monoisotopic or average."Conforms" without a number is not a measurement.
9Quantified contentMilligrams per vial, measured against a named reference standard.The commonest omission. This is the field that answers the question people think purity answers.
10Reference standardWhat the quantitation was calibrated against, with the standard’s own certificate referenced.Every quantitative number on the page inherits the uncertainty of an uncharacterised standard.
11Water contentKarl Fischer or loss on drying, as a percentage.You cannot reconcile gross vial mass with net peptide. Expect its absence at this tier.
12Counter-ion contentAcetate or trifluoroacetate, as a percentage by mass.As above. TFA can be several per cent of what you weighed.
13Residual solventsAcetonitrile, DMF, DCM by headspace GC, against ICH limits.Almost never present on research-grade documentation. Its absence is expected rather than suspicious.
14EndotoxinEU/mg by LAL or kinetic chromogenic method.Research material is not released against a pyrogen specification.
15SterilityGrowth or no growth in defined media over the full incubation period."Sterile filtered" is a process claim, not a result. The two are not equivalent and should not be read as such.
16Signature and laboratory identityA named laboratory with a verifiable report identifier.A document that cannot be traced to its issuer is a document you are trusting on typography.

Reconciling gross mass to label claim

This is where the purity-versus-content confusion becomes concrete. Suppose a vial is filled by mass to 11.62 mg of a lyophilised solid, and the certificate reports 99.0 per cent purity. How much peptide is in the vial?

StepValueWorking
Gross solid in the vial11.62 mgWhat was weighed at fill
Less residual water at 4.0%11.16 mg11.62 × 0.960
Less acetate counter-ion at 7.5%10.32 mg11.16 × 0.925
Less related impurities at 1.0%10.22 mg10.32 × 0.990
Net peptide10.22 mgAgainst a 10 mg label claim: +2.2%

That vial is in specification, and it is in specification because whoever filled it overfilled by 16 per cent to account for water and counter-ion. Now run the same arithmetic on a vial filled to 7.90 mg of solid at the same purity: net peptide 6.95 mg, against a 10 mg claim, a 30 per cent shortfall — with an identical purity figure on the certificate. The purity number cannot distinguish these two vials. Only a content assay can.

Why two competent laboratories disagree

They will, and by more than you expect. A spread of half a per cent to a per cent on the same vial is normal, and it is not evidence of incompetence on either side. The contributors, in rough order of magnitude:

CauseTypical effect on purityDirection
Detection wavelength (214 vs 280 nm)0.4–1.2 ppDepends on the aromatic content of the impurities
Gradient slope0.3–0.9 ppShallower resolves more impurities, so reports lower purity
Integration parameters and baseline0.1–0.5 ppEither
Column chemistry and age0.1–0.4 ppEither
Mobile-phase additive (TFA vs formic acid)0.1–0.4 ppEither; TFA usually sharpens peaks
Injection mass and detector linearity0.0–0.3 ppOverloading inflates apparent purity

The practical consequence: a gap of half a per cent between your result and the supplier's is noise. A gap of three per cent, reproduced on a second lot, is a finding. Set that threshold before you send the sample.

A short checklist you can actually use

  1. Photograph the vial and the certificate together, so the lot numbers are in one frame.
  2. Check the analysis date against the manufacturing date. Wrong order means wrong lot.
  3. Find the method section. If there is not one, downgrade everything below it.
  4. Find the detection wavelength. Note it, because you will need it to compare against your own result.
  5. Look for a content figure in milligrams. If it is absent, you have a purity certificate and no content claim.
  6. Look at the chromatogram, not the number. Count the impurity peaks.
  7. Write down, before you send anything for independent testing, what result would make you reject the material.
  8. Keep the certificate, the lot number and the invoice in one place. The dispute you might need them for is six months away.

Further reading on this site: the questions tagged coa749

coa

Certificates of analysis: what fields a useful one carries, how to tell a real analytical report from a marketing document, batch and lot…

749 questions purity788
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

788 questions
content-assay438
content-assay

Quantified content: how many milligrams of peptide are actually in the vial, measured against a calibrated reference standard. A separate test…

438 questions
and reference-standard179
reference-standard

Certified reference materials and in-house standards: traceability, the certificate of the standard itself, system suitability, and why every…

179 questions
.

Sponsored

PeptideMeter - Independent Peptide Analytics

Aggregated, published test results and vendor ratings built from submitted batches. Methodology stated, dataset browsable, no listing fees.

Browse results