Identity and purity are orthogonal, and a high purity says almost nothing about whether the peak is actually what you think it is.
Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.
More usefully, retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.
Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.
The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.
If you are ranking vendors, specify a method and have all samples tested at the same place.