Specifically, high-performance liquid chromatography is a separation technique, not an identification technique, and it answers the question "how much of the detected signal is a single species" not "which species is it."
Proline-rich sequences are particularly problematic because the isomerisation kinetics are in the same timescale as the separation, leading to split or broadened peaks at low temperature.
The relevant detail is that trifluoroacetic acid at 0.1 per cent is the standard ion-pairing agent because it suppresses tailing by neutralising the basic residues, but it also suppresses mass spectrometry signal.
Peptide separation by reverse-phase high-performance liquid chromatography is described in the general chapters of the United States Pharmacopeia, European Pharmacopeia and Japanese Pharmacopeia, and the methods converge on essentially the same principles.
I would caution against over-interpreting small differences between runs, because the noise floor of the method is larger than most people assume.
Ask for the chromatogram and the system suitability data, not just the number.