Read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.
On the detail: buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.