Accepted answer
1 percentage points, and the direction is the informative half. VendorInvestigate reports 99.4 per cent and the certificate 98.4, so the independent figure is higher. As impurity that is 0.6 per cent against 1.6 — 0.37 times as much unassigned area, which is the unusual direction and worth a second look at the integration. Method differences usually make the independent number the lower one, so an independent result above the certificate points at integration limits, a different wavelength, or a different definition of the main peak rather than at a better vial. Ask both parties for the method section before you decide which figure is wrong, because the answer is often neither.
The short answer is that two competent laboratories on identical material will disagree, and the disagreement is almost always explainable by method differences.
Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.
It helps to be literal here: column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.
The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.
Compare purity within a single laboratory on the same method, never across laboratories.