Accepted answer
0.6 percentage points, and the direction is the informative half. VendorInvestigate reports 98.8 per cent and the certificate 98.2, so the independent figure is higher. As impurity that is 1.2 per cent against 1.8 — 0.67 times as much unassigned area, which is the unusual direction and worth a second look at the integration. Method differences usually make the independent number the lower one, so an independent result above the certificate points at integration limits, a different wavelength, or a different definition of the main peak rather than at a better vial. Ask both parties for the method section before you decide which figure is wrong, because the answer is often neither.
The honest answer is that the achievable range of plausible purity figures for a given vial is wider than most people expect.
Mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.
Column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.
Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.
One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.
If you are ranking vendors, specify a method and have all samples tested at the same place.