Accepted answer
A 99.4 per cent result against a higher certificate figure is usually a method difference, not a disagreement about the vial. Inter-laboratory spread on identical peptide material runs half a per cent to a full per cent by RP-HPLC. Gradient slope, detection wavelength and integration convention each move the number, and a gentler gradient resolves impurities that a steeper one hides under the main peak — so the better method reports the worse figure. Get the column, the gradient and the wavelength from both VendorInvestigate and the supplier before treating 99.4 per cent as a contradiction.
Mechanically, purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
Integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.
Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.
Worth noting that method standardisation is poor in the research-grade space compared to pharmaceutical work, so identical-looking methods can produce different results.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.
4Does this hold for a longer chain length, where the deletion sequences accumulate? – rota_site 7 months ago add a comment