Accepted answer
The failure mode here is publishing a result that applies to the tested vial alone while implying it applies to the entire lot.
The sample size determination requires choosing a confidence level and an acceptable error rate, and the smaller the error rate you want, the larger your sample must be.
Reconciling gross mass to label claim
| Component | Typical share | Counted in purity? | Counted in content? |
|---|
| Target peptide | 88–94 % | Yes, as main peak | Yes |
| Related impurities | 1–3 % | Yes, as other peaks | No |
| Counter-ion (TFA or acetate) | 2–8 % | No | No |
| Residual water | 2–6 % | No | No |
| Bulking agent, if present | 0–40 % | No | No |
If the lot was manufactured in multiple batches, testing vials from each batch separately establishes whether batch-to-batch variation is acceptable.
Lyophilised peptide homogeneity studies show that vial-to-vial variation is usually small but occasionally large, depending on the distribution in the freeze-dryer.
One qualification: testing more vials gives better confidence, but at some point the cost outweighs the benefit.
If testing multiple vials, state how many you tested and why you chose those vials.
2Confirming from the other direction: I did the wrong thing and got exactly the predicted outcome. – hana_petrikova 3 months ago 3Is there a reason to prefer the second method over the first, other than cost? – rhian_prydderch 5 months ago add a comment