How do I reconcile 97.6% from Medutest with 95.2% from the supplier?

Asked 10 Feb 2026Modified 4 months agoViewed 5k times
1

Setup, so nobody has to ask: 97.6% · Medutest · 95.2%.

I noticed this today and I have not touched anything since, in case the state is diagnostic.

I have not discarded anything yet, so a test is still possible if that is the recommendation.

How do I distinguish the benign explanation from the one that matters?

purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

788 questions
batch-testing
batch-testing

Testing at the batch or lot level: sampling plans, how many vials from a lot need testing to say anything about the lot, and the difference…

865 questions
coa
coa

Certificates of analysis: what fields a useful one carries, how to tell a real analytical report from a marketing document, batch and lot…

749 questions
shareeditfollowflag
SF
askedsasha_ferreira15k1610 Feb 2026

3 Answers

Sorted by votes
24

Stated carefully, the honest answer is that the achievable range of plausible purity figures for a given vial is wider than most people expect.

Integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.

What each test answers

TestAnswersDoes NOT answer
RP-HPLC, area %What fraction of detected material is the targetHow much target is present
Quantified contentMilligrams of peptide per vialWhat the impurities are
ESI-MS identityWhether the molecular weight matchesPurity, or isomeric substitution
Peptide mappingSequence, localised to a fragmentQuantity
Karl FischerWater content of the solidSolvent content
LAL endotoxinPyrogen load in EU/mgSterility
Sterility testGrowth in defined media over 14 daysEndotoxin, or bioburden count

Mechanically, retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.

The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.

The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag
RP
answeredravi_pillai16k2814 Mar 2026
Sponsored

Sigma-Aldrich - Certified Reference Materials

Analytical standards and reagents with traceable certificates. Every quantitative result you read inherits the accuracy of the standard behind it.

Shop standards
16

Concretely, the short answer is that two competent laboratories on identical material will disagree, and the disagreement is almost always explainable by method differences.

Column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.

More usefully, tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.

The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
TA
answeredtess_amankwah48k3825 Mar 2026
2Worth adding that the method section is where the answer usually is. – tare_weight 6 months ago
add a comment
13

In practice, read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.

Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.

Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.

Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.

I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.

If you are ranking vendors, specify a method and have all samples tested at the same place.

edited 8 Apr 2026 by triple_agonist_q — added the method parameters

shareimprove this answerflag
TQ
answeredtriple_agonist_q37k385 Apr 2026
8I tested this on two lots and got the same answer, so at least it reproduces. – Dr_Elias_Weiss 8 months ago
The timing signature is the useful part. Everything else is confounded. – Dr_Bram_Verhoeven 9 months ago