Accepted answer
Specifically, purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.
Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
If you are ranking vendors, specify a method and have all samples tested at the same place.