Accepted answer
Mechanically, the method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
Reconciling gross mass to label claim
| Component | Typical share | Counted in purity? | Counted in content? |
|---|
| Target peptide | 88–94 % | Yes, as main peak | Yes |
| Related impurities | 1–3 % | Yes, as other peaks | No |
| Counter-ion (TFA or acetate) | 2–8 % | No | No |
| Residual water | 2–6 % | No | No |
| Bulking agent, if present | 0–40 % | No | No |
In practice, retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.
Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.