Accepted answer
On the detail: gradient slope is the most powerful parameter and almost nobody mentions it, which is why two reports on the same material disagree by a point.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
Concretely, buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.
Compare purity within a single laboratory on the same method, never across laboratories.