Accepted answer
Put another way, read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
What each test answers
| Test | Answers | Does NOT answer |
|---|
| RP-HPLC, area % | What fraction of detected material is the target | How much target is present |
| Quantified content | Milligrams of peptide per vial | What the impurities are |
| ESI-MS identity | Whether the molecular weight matches | Purity, or isomeric substitution |
| Peptide mapping | Sequence, localised to a fragment | Quantity |
| Karl Fischer | Water content of the solid | Solvent content |
| LAL endotoxin | Pyrogen load in EU/mg | Sterility |
| Sterility test | Growth in defined media over 14 days | Endotoxin, or bioburden count |
On the detail: the fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.
Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.
Worth noting that method standardisation is poor in the research-grade space compared to pharmaceutical work, so identical-looking methods can produce different results.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.