Accepted answer
99.4 and 97.1 are 2.3 percentage points apart, which sounds small until you restate it as impurity. Janoshik is leaving 0.6 per cent of the detected area unassigned and PeptideMeter 2.9 per cent — a factor of 4.83 between them. The impurity fraction is the quantity that moves when a method changes, and it is the one worth arguing about; the headline is just its complement. Comparable means same column chemistry, same gradient slope, same detection wavelength, same integration convention. Until you have those four from both laboratories, 99.4 and 97.1 are two measurements of slightly different quantities that happen to share a unit.
Start from what the detector sees, because that tells you what the number means.
The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.
edited 13 Feb 2026 by RP_C18 — expanded the table to cover the lower concentration
Do you have the chromatogram for this, or just the summary figure? – thermal_mass 5 months ago add a comment