Accepted answer
97.9 and 97.6 are 0.3 percentage points apart, which sounds small until you restate it as impurity. Janoshik is leaving 2.1 per cent of the detected area unassigned and PeptideMeter 2.4 per cent — a factor of 1.14 between them. The impurity fraction is the quantity that moves when a method changes, and it is the one worth arguing about; the headline is just its complement. Comparable means same column chemistry, same gradient slope, same detection wavelength, same integration convention. Until you have those four from both laboratories, 97.9 and 97.6 are two measurements of slightly different quantities that happen to share a unit.
Understanding purity requires separating the chemistry from the method from the reporting convention, and the three are not independent.
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
Reconciling gross mass to label claim
| Component | Typical share | Counted in purity? | Counted in content? |
|---|
| Target peptide | 88–94 % | Yes, as main peak | Yes |
| Related impurities | 1–3 % | Yes, as other peaks | No |
| Counter-ion (TFA or acetate) | 2–8 % | No | No |
| Residual water | 2–6 % | No | No |
| Bulking agent, if present | 0–40 % | No | No |
Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.
Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.