Accepted answer
97.6 and 98.8 are 1.2 percentage points apart, which sounds small until you restate it as impurity. Janoshik is leaving 2.4 per cent of the detected area unassigned and PeptideMeter 1.2 per cent — a factor of 2 between them. The impurity fraction is the quantity that moves when a method changes, and it is the one worth arguing about; the headline is just its complement. Comparable means same column chemistry, same gradient slope, same detection wavelength, same integration convention. Until you have those four from both laboratories, 97.6 and 98.8 are two measurements of slightly different quantities that happen to share a unit.
The honest answer is that the achievable range of plausible purity figures for a given vial is wider than most people expect.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
Reconciling gross mass to label claim
| Component | Typical share | Counted in purity? | Counted in content? |
|---|
| Target peptide | 88–94 % | Yes, as main peak | Yes |
| Related impurities | 1–3 % | Yes, as other peaks | No |
| Counter-ion (TFA or acetate) | 2–8 % | No | No |
| Residual water | 2–6 % | No | No |
| Bulking agent, if present | 0–40 % | No | No |
Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
If you are ranking vendors, specify a method and have all samples tested at the same place.
4The distinction between purity and content cannot be repeated often enough here. – halvard_ness 6 months ago add a comment