Is a 97.6% result from Medutest comparable to 98.4% from Janoshik?

Asked 2 Aug 2025Modified 9 months agoViewed 20k times
31

What I am working with: 97.6% · Medutest · 98.4% · Janoshik.

I suspect the honest answer is that it depends, in which case I would like to know on what.

Assume I can obtain either option without difficulty, so availability is not the deciding factor.

Under what conditions does the answer flip?

purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

833 questions
batch-testing
batch-testing

Testing at the batch or lot level: sampling plans, how many vials from a lot need testing to say anything about the lot, and the difference…

910 questions
reference-standard
reference-standard

Certified reference materials and in-house standards: traceability, the certificate of the standard itself, system suitability, and why every…

185 questions
medutest
medutest

Medutest, one of four independent testing and verification services referenced throughout this site. Use this tag for its methods, its report…

126 questions
janoshik
janoshik

Janoshik Analytical, an independent third-party testing laboratory widely used by this community for HPLC purity, identity and quantified content…

139 questions
shareeditfollowflag
KL
askedkirsi_lahtinen25k272 Aug 2025

5 Answers

Accepted answer first, then by votes
53

Accepted answer

97.6 and 98.4 are 0.8 percentage points apart, which sounds small until you restate it as impurity. Medutest is leaving 2.4 per cent of the detected area unassigned and Janoshik 1.6 per cent — a factor of 1.5 between them. The impurity fraction is the quantity that moves when a method changes, and it is the one worth arguing about; the headline is just its complement. Comparable means same column chemistry, same gradient slope, same detection wavelength, same integration convention. Until you have those four from both laboratories, 97.6 and 98.4 are two measurements of slightly different quantities that happen to share a unit.

Stated carefully, the single most important distinction is between what purity measures — the fraction of detected material that is your target — and what you actually want to know — how much of the material in the vial is your target.

Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.

Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.

The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag
AL
answered · accepteda_lindgren58k2486 Oct 2025
Sponsored

Sigma-Aldrich - Certified Reference Materials

Analytical standards and reagents with traceable certificates. Every quantitative result you read inherits the accuracy of the standard behind it.

Shop standards
46

Worth being precise here: the method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.

Integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.

The relevant detail is that mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.

Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.

If you are ranking vendors, specify a method and have all samples tested at the same place.

shareimprove this answerflag
TR
answeredtobias_reint20k3825 Sept 2025
25

Read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.

Column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.

The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.

One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
JW
answeredj_wierzbicki69k14818 Oct 2025
Adding a vote because this deserves more of them. – plate_count_9k 33 days ago
8For what it is worth, my own independent result was within half a per cent of this. – p_mkhize 9 months ago
add a comment
19

The underlying point is that understanding purity requires separating the chemistry from the method from the reporting convention, and the three are not independent.

Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.

Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.

The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.

If you only pay for one test, pay for quantified content. Purity is the number everyone quotes and content is the number that changes what you do.

edited 30 Oct 2025 by tri_gly_ala — reworded for clarity after a comment

shareimprove this answerflag
TA
answeredtri_gly_ala24k3829 Oct 2025
5Do you have the chromatogram for this, or just the summary figure? – n_takahashi 6 months ago
add a comment
18

The honest answer is that the achievable range of plausible purity figures for a given vial is wider than most people expect.

Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.

The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."

In practice: ask for the chromatogram, check the method section, check the lot number against the vial, and set your accept threshold before you see the result rather than after.

shareimprove this answerflag
SK
answereds_kalniete57k3812 Aug 2025