Accepted answer
More usefully, the short answer is that two competent laboratories on identical material will disagree, and the disagreement is almost always explainable by method differences.
Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
Worth being precise here: column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.
The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.
One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.
Compare purity within a single laboratory on the same method, never across laboratories.
Any reason this would differ for a longer peptide? – dead_volume 7 months ago add a comment