In practice, the single most important distinction is between what purity measures — the fraction of detected material that is your target — and what you actually want to know — how much of the material in the vial is your target.
Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
Reconciling gross mass to label claim
| Component | Typical share | Counted in purity? | Counted in content? |
|---|
| Target peptide | 88–94 % | Yes, as main peak | Yes |
| Related impurities | 1–3 % | Yes, as other peaks | No |
| Counter-ion (TFA or acetate) | 2–8 % | No | No |
| Residual water | 2–6 % | No | No |
| Bulking agent, if present | 0–40 % | No | No |
Column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.
The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."
Compare purity within a single laboratory on the same method, never across laboratories.