System suitability is the part of a report that tells you whether the method was under control on the day you were tested, and its absence is concerning.
Mobile phase pH at the point where you inject must match the mobile phase pH at the start of the gradient, or the sample will not be focused at the column head.
Peptide separation by reverse-phase high-performance liquid chromatography is described in the general chapters of the United States Pharmacopeia, European Pharmacopeia and Japanese Pharmacopeia, and the methods converge on essentially the same principles.
One qualification: the limit of detection on a reversed-phase HPLC is set by the noise on the baseline, which for these molecules is usually in the range of a tenth of one per cent or less, and anything smaller is not reproducibly detectable.
If you only pay for one test, pay for quantified content. Purity is the number everyone quotes and content is the number that changes what you do.
edited 21 Aug 2024 by grainne_ahearn — reworded for clarity after a comment
5The system-suitability data is the part that tells you whether to believe the rest. – gradient_slope 9 months ago add a comment