The underlying point is that read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.
4Worth adding that the method section is where the answer usually is. – Dr_Aoife_Brennan 10 months ago add a comment