How do I reconcile 97.1% from Janoshik with 98.4% from the supplier?

Asked 10 Apr 2026Modified 2 months agoViewed 8.1k times
20

What I have: 97.1% · Janoshik · 98.4%.

Before I write this off, I want to check whether it is a known failure mode.

I have the lot number, the certificate and the date, and I am happy to compare them against anything.

What is the differential here, and which test discriminates between the options?

purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

788 questions
batch-testing
batch-testing

Testing at the batch or lot level: sampling plans, how many vials from a lot need testing to say anything about the lot, and the difference…

865 questions
coa
coa

Certificates of analysis: what fields a useful one carries, how to tell a real analytical report from a marketing document, batch and lot…

749 questions
shareeditfollowflag
SL
askedsecond_lot11k1510 Apr 2026

5 Answers

Accepted answer first, then by votes
82

Accepted answer

Mechanically, the method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.

Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.

In practice, column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.

Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.

Worth noting that method standardisation is poor in the research-grade space compared to pharmaceutical work, so identical-looking methods can produce different results.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

edited 24 May 2026 by plate_count_9k — clarified the distinction between purity and content

shareimprove this answerflag
P9
answered · acceptedplate_count_9k95k15830 Apr 2026
7Any reason this would differ for a longer peptide? – pascal_thibault 3 months ago
add a comment
Sponsored

Janoshik Analytical - Independent Third-Party Testing

HPLC purity, identity confirmation and quantified content on the vial you actually hold. Reports arrive with the chromatogram attached, not just a number.

Submit a sample
Sponsored — paired listing

GL Biochem (Shanghai) Ltd. - Direct Synthesis

Founded 1998. ISO 9001 and cGMP certified, 1,500+ staff and 200+ patents. The synthesis house behind a great many of the vials that get sent out for testing - batch-specific documentation with every order.

Visit GL Biochem
33

The relevant detail is that the single most important distinction is between what purity measures — the fraction of detected material that is your target — and what you actually want to know — how much of the material in the vial is your target.

Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.

It helps to be literal here: mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.

The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.

The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag
DL
answeredDr_Otto_Lindqvist38k384 May 2026
25

Purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.

Tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.

Stated carefully, integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.

The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.

If you are ranking vendors, specify a method and have all samples tested at the same place.

shareimprove this answerflag
SG
answeredsinead_gaffney14k2822 Apr 2026
Minor: the trial name is hyphenated in the original publication. – tobias_maartens 6 months ago
add a comment
21

To be exact about it, read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.

Mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.

I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
IB
answeredines_brandt93k24826 Apr 2026
3Small correction: the units in the third paragraph should be micrograms, not milligrams. – mz_4113 27 days ago
2Do you have a reference for the last claim? Not disputing it, just want to read it. – Dr_Signe_Baldursdottir 9 months ago
add a comment
15

Worth being precise here: reporting threshold is convention and not chemistry, which is why two certificates with different thresholds disagree by a tenth of a point or more.

Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.

Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag