To be exact about it, gradient slope is the most powerful parameter and almost nobody mentions it, which is why two reports on the same material disagree by a point.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
Integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.
The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
Compare purity within a single laboratory on the same method, never across laboratories.
edited 4 Jan 2026 by h_villanueva — removed a claim I could not source