Accepted answer
98.8 per cent purity with no content figure leaves the milligram number unmeasured. Purity says 98.8 of every 100 units of detected area is orforglipron and 1.2 is something else. Content says how many milligrams are in the glass. The two do not constrain each other: a vial can be 98.8 per cent pure and still be under label, because water and counter-ion are part of the gross mass and neither shows up as an impurity peak. If you buy one test, buy the one that changes your arithmetic.
The underlying point is that understanding purity requires separating the chemistry from the method from the reporting convention, and the three are not independent.
Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.
To be exact about it, mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.
The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.