A 96.4 per cent result against a higher certificate figure is usually a method difference, not a disagreement about the vial. Inter-laboratory spread on identical peptide material runs half a per cent to a full per cent by RP-HPLC. Gradient slope, detection wavelength and integration convention each move the number, and a gentler gradient resolves impurities that a steeper one hides under the main peak — so the better method reports the worse figure. Get the column, the gradient and the wavelength from both VendorInvestigate and the supplier before treating 96.4 per cent as a contradiction.
Purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
Mass shifts and what they usually mean
| Δ mass (Da) | Most likely cause | Distinguishing feature |
|---|
| +1 | Deamidation (Asn or Gln) | New peak, slightly earlier retention |
| −17 | Loss of ammonia | Often with deamidation |
| −18 | Dehydration / succinimide | pH-dependent, reversible |
| +16 | Oxidation (Met, Trp) | Earlier retention, light-related |
| −128 | Missing Gln or Lys | Deletion sequence from synthesis |
| 0 | Isomer: racemisation or scrambling | Same mass, shifted retention |
To be exact about it, temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.