Purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.
The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."
If you are ranking vendors, specify a method and have all samples tested at the same place.
7Do you have the chromatogram for this, or just the summary figure? – k_szabo 8 months ago 6Any reason to prefer ion chromatography over fluorine NMR for the counter-ion here? – two_two_micron 6 months ago add a comment