To be exact about it, reverse-phase HPLC is the workhorse for peptide purity work, but it is almost universally run under conditions that are not optimal for a peptide of this chain length.
Proline-rich sequences are particularly problematic because the isomerisation kinetics are in the same timescale as the separation, leading to split or broadened peaks at low temperature.
The resolving power of a separation is quantified by the resolution parameter R, defined from the heights and widths of adjacent peaks, and pharmacopoeial methods typically demand R greater than 1.5 for a method to be considered validated.
One qualification: the limit of detection on a reversed-phase HPLC is set by the noise on the baseline, which for these molecules is usually in the range of a tenth of one per cent or less, and anything smaller is not reproducibly detectable.
In practice: ask for the chromatogram, check the method section, check the lot number against the vial, and set your accept threshold before you see the result rather than after.