High-performance liquid chromatography is a separation technique, not an identification technique, and it answers the question "how much of the detected signal is a single species" not "which species is it."
The 214 nanometre wavelength is chosen because it corresponds to the amide backbone absorption, making response roughly proportional to the number of peptide bonds.
The part that matters: wider-pore phases — 300 Angstrom rather than 100 Angstrom — have faster mass transfer and narrower peaks for peptides above three kilodaltons, which is almost every peptide you will see.
Worth noting that the achievable resolution depends on the chemistry of the molecule — some peptide sequences separate easily while others are notoriously difficult regardless of method.
Ask for the chromatogram and the system suitability data, not just the number.