Accepted answer
Specifically, freeze-thaw damage happens at the moving ice front, not at the storage temperature. Once the sample is frozen solid and cold, very little is happening. The damage is done during freezing and thawing, which is why the number of cycles matters and the duration of the hold mostly does not.
Aggregation is the failure mode that reverse-phase HPLC is worst at detecting, because a large soluble aggregate may not elute at all and an insoluble one is filtered out during sample preparation. If your purity result comes back normal but the vial looks hazy, believe the vial. Size-exclusion chromatography is the method that sees this.
Degradation pathway by condition
| Pathway | Dominant when | Detected by |
|---|
| Deamidation | Solution, neutral to alkaline pH | RP-HPLC, +1 Da on MS |
| Oxidation | Light, trace metals, peroxides | RP-HPLC, +16 Da on MS |
| Hydrolysis | Solution, extremes of pH | RP-HPLC, fragment masses |
| Aggregation | Agitation, interfaces, high concentration | SEC, visual haze; often invisible on RP-HPLC |
| Freeze-concentration damage | Freeze-thaw of buffered solution | SEC, loss of recovered content |
Light matters for specific residues rather than in general. Tryptophan and to a lesser extent tyrosine and methionine are photo-labile; a sequence without them is largely indifferent to ambient light over the timescales in question. Amber glass is cheap insurance rather than a requirement.
Deamidation kinetics for asparagine in peptides are well characterised and strongly sequence-dependent: the residue following the asparagine dominates the rate, with glycine and serine at the n+1 position accelerating it by an order of magnitude relative to bulkier residues. That is why two peptides in the same buffer at the same temperature can have quite different shelf lives.
Store solid, store cold, store dry, and reconstitute what you will use rather than what fits in the vial.