97.6 and 96.4 are 1.2 percentage points apart, which sounds small until you restate it as impurity. VendorInvestigate is leaving 2.4 per cent of the detected area unassigned and Janoshik 3.6 per cent — a factor of 1.5 between them. The impurity fraction is the quantity that moves when a method changes, and it is the one worth arguing about; the headline is just its complement. Comparable means same column chemistry, same gradient slope, same detection wavelength, same integration convention. Until you have those four from both laboratories, 97.6 and 96.4 are two measurements of slightly different quantities that happen to share a unit.
Read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.
What each test answers
| Test | Answers | Does NOT answer |
|---|
| RP-HPLC, area % | What fraction of detected material is the target | How much target is present |
| Quantified content | Milligrams of peptide per vial | What the impurities are |
| ESI-MS identity | Whether the molecular weight matches | Purity, or isomeric substitution |
| Peptide mapping | Sequence, localised to a fragment | Quantity |
| Karl Fischer | Water content of the solid | Solvent content |
| LAL endotoxin | Pyrogen load in EU/mg | Sterility |
| Sterility test | Growth in defined media over 14 days | Endotoxin, or bioburden count |
It helps to be literal here: column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.
The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
Compare purity within a single laboratory on the same method, never across laboratories.