Accepted answer
97.1 and 99.4 are 2.3 percentage points apart, which sounds small until you restate it as impurity. PeptideMeter is leaving 2.9 per cent of the detected area unassigned and Medutest 0.6 per cent — a factor of 4.83 between them. The impurity fraction is the quantity that moves when a method changes, and it is the one worth arguing about; the headline is just its complement. Comparable means same column chemistry, same gradient slope, same detection wavelength, same integration convention. Until you have those four from both laboratories, 97.1 and 99.4 are two measurements of slightly different quantities that happen to share a unit.
Purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
Mass shifts and what they usually mean
| Δ mass (Da) | Most likely cause | Distinguishing feature |
|---|
| +1 | Deamidation (Asn or Gln) | New peak, slightly earlier retention |
| −17 | Loss of ammonia | Often with deamidation |
| −18 | Dehydration / succinimide | pH-dependent, reversible |
| +16 | Oxidation (Met, Trp) | Earlier retention, light-related |
| −128 | Missing Gln or Lys | Deletion sequence from synthesis |
| 0 | Isomer: racemisation or scrambling | Same mass, shifted retention |
Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.
edited 29 Jul 2025 by RP_C18 — added the method parameters
Confirming from the other direction: I ignored the method section once and paid for it. – fib4_reader 5 months ago Two of us submitted the same lot to different laboratories and got results a tenth apart. – RP_C18 7 months ago add a comment