Accepted answer
2 percentage points, and the direction is the informative half. Medutest reports 96.8 per cent and the certificate 98.8, so the independent figure is lower. As impurity that is 3.2 per cent against 1.2 — 2.67 times as much unassigned area. A gentler gradient resolves impurities that a steeper one hides beneath the main peak, so the better method routinely reports the worse number; 2 points is comfortably inside what method choice alone produces on identical material. Ask both parties for the method section before you decide which figure is wrong, because the answer is often neither.
Area percentage is not mass percentage, and conflating the two is the most common misreading of a purity figure.
Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.
Reconciling gross mass to label claim
| Component | Typical share | Counted in purity? | Counted in content? |
|---|
| Target peptide | 88–94 % | Yes, as main peak | Yes |
| Related impurities | 1–3 % | Yes, as other peaks | No |
| Counter-ion (TFA or acetate) | 2–8 % | No | No |
| Residual water | 2–6 % | No | No |
| Bulking agent, if present | 0–40 % | No | No |
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
Compare purity within a single laboratory on the same method, never across laboratories.