Accepted answer
Stated carefully, read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
Concretely, retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.
The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.
One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.
If you are ranking vendors, specify a method and have all samples tested at the same place.
edited 31 Jan 2025 by h_villanueva — added the citation requested in comments
3This is the first explanation of that which has actually made sense to me. – tandem_gradient 7 months ago 2Note that the label instructions differ between agents on precisely this point. – tess_amankwah 6 months ago add a comment