On the detail: read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.
Put another way, buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.
Worth noting that method standardisation is poor in the research-grade space compared to pharmaceutical work, so identical-looking methods can produce different results.
Compare purity within a single laboratory on the same method, never across laboratories.