Accepted answer
To be exact about it, understanding purity requires separating the chemistry from the method from the reporting convention, and the three are not independent.
Integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.
What each test answers
| Test | Answers | Does NOT answer |
|---|
| RP-HPLC, area % | What fraction of detected material is the target | How much target is present |
| Quantified content | Milligrams of peptide per vial | What the impurities are |
| ESI-MS identity | Whether the molecular weight matches | Purity, or isomeric substitution |
| Peptide mapping | Sequence, localised to a fragment | Quantity |
| Karl Fischer | Water content of the solid | Solvent content |
| LAL endotoxin | Pyrogen load in EU/mg | Sterility |
| Sterility test | Growth in defined media over 14 days | Endotoxin, or bioburden count |
The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.
The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.
The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."
Compare purity within a single laboratory on the same method, never across laboratories.