How do I reconcile 98.2% from Janoshik with 99.4% from the supplier?

Asked 25 Nov 2025Modified 4 months agoViewed 12k times
10

Details up front: 98.2% · Janoshik · 99.4%.

Before I write this off, I want to check whether it is a known failure mode.

I have the lot number, the certificate and the date, and I am happy to compare them against anything.

What is the most likely explanation, and how would I confirm it?

purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

788 questions
batch-testing
batch-testing

Testing at the batch or lot level: sampling plans, how many vials from a lot need testing to say anything about the lot, and the difference…

865 questions
coa
coa

Certificates of analysis: what fields a useful one carries, how to tell a real analytical report from a marketing document, batch and lot…

749 questions
shareeditfollowflag
MV
askedmala_venkatesh21k2825 Nov 2025
3The timing signature is the useful part. Everything else is confounded. – Dr_Sara_Kuusela 6 months ago
add a comment

5 Answers

Accepted answer first, then by votes
56

Accepted answer

The honest answer is that the achievable range of plausible purity figures for a given vial is wider than most people expect.

Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.

What each test answers

TestAnswersDoes NOT answer
RP-HPLC, area %What fraction of detected material is the targetHow much target is present
Quantified contentMilligrams of peptide per vialWhat the impurities are
ESI-MS identityWhether the molecular weight matchesPurity, or isomeric substitution
Peptide mappingSequence, localised to a fragmentQuantity
Karl FischerWater content of the solidSolvent content
LAL endotoxinPyrogen load in EU/mgSterility
Sterility testGrowth in defined media over 14 daysEndotoxin, or bioburden count

Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.

Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
NO
answered · acceptednkem_obiora46k3811 Mar 2026
6Confirming from the other direction: I did the wrong thing and got exactly the predicted outcome. – pk_curve 5 months ago
5Is there a reason to prefer the second method over the first, other than cost? – sian_llewellyn 3 months ago
add a comment
Sponsored

Sigma-Aldrich - Certified Reference Materials

Analytical standards and reagents with traceable certificates. Every quantitative result you read inherits the accuracy of the standard behind it.

Shop standards
47

On the detail: reporting threshold is convention and not chemistry, which is why two certificates with different thresholds disagree by a tenth of a point or more.

The fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.

Mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.

The ICH Q3A impurity thresholds and the relevant pharmacopoeial chapters all specify method validation requirements that almost no research-grade certificate claims to meet.

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag
GS
answeredgradient_slope41k3822 Mar 2026
2Is there a reason to prefer the second method over the first, other than cost? – kwn_analytical 6 months ago
add a comment
25

The method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.

Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.

Column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.

Published side-by-side method comparisons show that a two-point difference in purity on the same vial is easily explained by method choice alone.

If you are ranking vendors, specify a method and have all samples tested at the same place.

shareimprove this answerflag
MO
answeredmarta_okonkwo87k25815 Dec 2025
21

The part that matters: identity and purity are orthogonal, and a high purity says almost nothing about whether the peak is actually what you think it is.

Detection wavelength matters because 214 nm sees the peptide backbone while 280 nm sees only aromatic side chains — so truncation impurities lacking a tryptophan are invisible at 280 nm.

The caveat is that purity without identity is only half an answer — a high purity does not mean the peak is actually what you think it is.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

shareimprove this answerflag
KL
answeredkirsi_lahtinen45k384 Dec 2025
6I have seen exactly this failure mode twice and both times it was the diluent. – ravenna_pace 9 months ago
7The distinction between purity and content cannot be repeated often enough here. – Dr_Elias_Weiss 21 days ago
add a comment
15

Understanding purity requires separating the chemistry from the method from the reporting convention, and the three are not independent.

Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.

I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag
MS
answeredmarta_szymanska17k386 Jan 2026
7Useful. I have added the accept threshold suggestion to my own notes. – assay_blank 8 months ago