Specifically, purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.
Column pore size affects mass transfer — a 100 Angstrom packing on a 5 kDa peptide restricts diffusion, broadening the peak and potentially hiding small impurities in the shoulders.
Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.
edited 9 Nov 2025 by Dr_Lena_Ostrowska — added the method parameters