How do I reconcile 97.6% from Janoshik with 98.8% from the supplier?

Asked 17 Jan 2025Modified 15 months agoViewed 28k times
29

Numbers first: 97.6% · Janoshik · 98.8%.

Before I write this off, I want to check whether it is a known failure mode.

I have the lot number, the certificate and the date, and I am happy to compare them against anything.

What is the differential here, and which test discriminates between the options?

purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

788 questions
batch-testing
batch-testing

Testing at the batch or lot level: sampling plans, how many vials from a lot need testing to say anything about the lot, and the difference…

865 questions
coa
coa

Certificates of analysis: what fields a useful one carries, how to tell a real analytical report from a marketing document, batch and lot…

749 questions
shareeditfollowflag
CL
askedcap_the_luer15k2817 Jan 2025

5 Answers

Accepted answer first, then by votes
72

Accepted answer

Start from what the detector sees, because that tells you what the number means.

Sample solvent strength affects peak shape — if you inject in strong solvent on a gradient starting in weak solvent, the solvent peak can distort your main peak or create a false shoulder.

What each test answers

TestAnswersDoes NOT answer
RP-HPLC, area %What fraction of detected material is the targetHow much target is present
Quantified contentMilligrams of peptide per vialWhat the impurities are
ESI-MS identityWhether the molecular weight matchesPurity, or isomeric substitution
Peptide mappingSequence, localised to a fragmentQuantity
Karl FischerWater content of the solidSolvent content
LAL endotoxinPyrogen load in EU/mgSterility
Sterility testGrowth in defined media over 14 daysEndotoxin, or bioburden count

More usefully, the fraction of your main peak that is actually your target versus isomers, fragments or related sequences is invisible without complementary identity testing.

Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.

Worth noting that method standardisation is poor in the research-grade space compared to pharmaceutical work, so identical-looking methods can produce different results.

If you are ranking vendors, specify a method and have all samples tested at the same place.

shareimprove this answerflag
TQ
answered · acceptedtriple_agonist_q37k383 Feb 2025
Sponsored

PeptideMeter - Independent Peptide Analytics

Aggregated, published test results and vendor ratings built from submitted batches. Methodology stated, dataset browsable, no listing fees.

Browse results
61

The relevant detail is that the method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.

Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.

Mass on column affects detector linearity and peak overlap — overloading broadens peaks and hides neighbours, while underloading improves resolution but loses sensitivity.

The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.

The limitation is that single-digit micro-impurities become invisible at typical reporting thresholds, so "no impurities detected" means "none above one in two thousand."

Compare purity within a single laboratory on the same method, never across laboratories.

shareimprove this answerflag
DV
answeredDr_Bram_Verhoeven85k24814 Feb 2025
5Related: the same reasoning applies to the counter-ion question. – lipid_panel_q 7 months ago
add a comment
32

Put another way, understanding purity requires separating the chemistry from the method from the reporting convention, and the three are not independent.

Gradient slope controls resolution, and gentler slopes resolve co-eluting impurities into separate peaks — so the better method reports the worse purity number.

Integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.

I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.

The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.

edited 14 Mar 2025 by ellis_thorne — corrected a unit error in the worked example

shareimprove this answerflag
ET
answeredellis_thorne17k179 Mar 2025
The arithmetic checks out. I ran the same numbers and got the same result. – tenth_of_a_unit 6 months ago
add a comment
27

The part that matters: read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.

Buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.

If you are ranking vendors, specify a method and have all samples tested at the same place.

shareimprove this answerflag
ED
answerede_dziedzic87k24826 Feb 2025
3For what it is worth, my own result was within half a per cent of this. – meniscus_film 10 months ago
4Any reason this would differ for a longer peptide? – coldpack_88 41 days ago
add a comment
19

The underlying point is that reporting threshold is convention and not chemistry, which is why two certificates with different thresholds disagree by a tenth of a point or more.

Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.

One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.

Compare purity within a single laboratory on the same method, never across laboratories.

edited 29 Apr 2025 by Dr_Malik_Osei — clarified the distinction between purity and content

shareimprove this answerflag
DO
answeredDr_Malik_Osei37k3831 Mar 2025