Accepted answer
Specifically, purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Retention time is sequence-specific and method-specific, so comparing your result to a supplier value using a different method is meaningless without method documentation.
Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
Inter-laboratory studies on identical peptide material routinely find half-a-per-cent to a full-per-cent spreads in reported purity on the same sample.
I would be careful about over-reading a single measurement — treat it as a data point, not as ground truth.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.