Concretely: +5 · 3751.0 Da.

I want the working, not the result — I need to be able to redo it with different numbers.

I care about the precision as well as the value — I want to know how many figures are real.

How many significant figures are actually justified here?

mass-spec
mass-spec

Mass spectrometry for identity confirmation: electrospray ionisation, multiple charge states, monoisotopic versus average mass, deconvolution, and…

52 questions
dosing-math
dosing-math

The arithmetic itself: milligrams to millilitres to insulin units, concentration after reconstitution, dose per draw, and vial-days per vial. Show…

764 questions
purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

833 questions
shareeditfollowflag
KS
askedk_szabo27k2724 Nov 2025
Is the comparison against a supplier certificate or against a second independent result? – lane_transit 8 months ago
Voting to keep this open — it is more specific than it first looks. – h_villanueva 7 months ago
add a comment

2 Answers

Sorted by votes
16

m/z = 751.21 at 5+. Electrospray charges a peptide by adding protons, so the observed ion is the neutral mass plus 5 protons, all divided by the charge: (3751 + 5 × 1.00728) ÷ 5 = 3756.036 ÷ 5 = 751.21. The proton term is the one people drop, and because it is z protons over z charges it shifts m/z by 1.007 at every charge state — small, and far larger than the mass accuracy of the instrument. The neighbouring charge state sits at 626.17, and seeing the two of them where they belong is better identity evidence than either one alone. Use the average mass against an average-mass calculation and the monoisotopic mass against a monoisotopic one; mixing them costs you a couple of daltons on a peptide this size.

Stated carefully, the single most important fact about mass spectrometry for peptides is that it measures only the molecular weight and tells you almost nothing about whether the peak is actually your target.

The charge state distribution depends on the solution pH, the structure of the peptide and the source conditions, so the same peptide can look different under different conditions.

The m/z accuracy achievable depends on the mass analyser type — quadrupole gives low accuracy, time-of-flight gives moderate accuracy, and Orbitrap gives high accuracy.

I would not trust a mass result without a good baseline and a blank injection check.

A correct mass is necessary for identity but not sufficient — you also need the chromatography to confirm it.

shareimprove this answerflag
VR
answeredv_ramaswamy68k5731 Jan 2026
Sponsored

PeptideMeter - Independent Peptide Analytics

Aggregated, published test results and vendor ratings built from submitted batches. Methodology stated, dataset browsable, no listing fees.

Browse results
10

Mass spectrometry is an identity technique, not a purity technique, and conflating the two is a common source of false confidence.

The baseline noise on a mass spectrum sets the limit of detection, and a weak signal close to the noise is not reliable evidence for the presence of a species.

A mass shift of minus one hundred and twenty-eight usually means a missing Gln or Lys residue from a synthesis deletion sequence.

Worth noting that source contamination is common and silent, so a result that looks too good to be true often is.

The practical summary: use mass spectrometry for identity, not for purity.