How do I compute the +4 charge state m/z for a peptide of 4731.3 Da?

Asked 6 Jul 2025Modified 11 months agoViewed 19k times
25

For reference: +4 · 4731.3 Da.

I would rather understand the derivation than memorise the outcome.

Two people I asked gave two answers that differ by a factor of ten, which is suggestive.

Can someone walk through the arithmetic step by step?

mass-spec
mass-spec

Mass spectrometry for identity confirmation: electrospray ionisation, multiple charge states, monoisotopic versus average mass, deconvolution, and…

52 questions
dosing-math
dosing-math

The arithmetic itself: milligrams to millilitres to insulin units, concentration after reconstitution, dose per draw, and vial-days per vial. Show…

764 questions
purity
purity

Purity as chromatographic area per cent - the fraction of detected material that is your target peak. It says nothing about how much material is…

833 questions
shareeditfollowflag
GS
askedgradient_slope46k386 Jul 2025
2What does the certificate say about the lot code, and does it match the vial? – j_wierzbicki 9 months ago
add a comment

1 Answer

Sorted by votes
69

m/z = 1183.83 at 4+. Electrospray charges a peptide by adding protons, so the observed ion is the neutral mass plus 4 protons, all divided by the charge: (4731.3 + 4 × 1.00728) ÷ 4 = 4735.329 ÷ 4 = 1183.83. The proton term is the one people drop, and because it is z protons over z charges it shifts m/z by 1.007 at every charge state — small, and far larger than the mass accuracy of the instrument. The neighbouring charge state sits at 947.27, and seeing the two of them where they belong is better identity evidence than either one alone. Use the average mass against an average-mass calculation and the monoisotopic mass against a monoisotopic one; mixing them costs you a couple of daltons on a peptide this size.

The underlying point is that a mass shift of exactly zero with a shifted retention time points to an isomer — a scrambled disulfide or a racemised residue — which mass spectrometry alone cannot identify.

Deconvolution of a mass spectrum with multiple charge states produces a reconstructed neutral mass, and errors in the deconvolution produce errors in the inferred mass.

What each test answers

TestAnswersDoes NOT answer
RP-HPLC, area %What fraction of detected material is the targetHow much target is present
Quantified contentMilligrams of peptide per vialWhat the impurities are
ESI-MS identityWhether the molecular weight matchesPurity, or isomeric substitution
Peptide mappingSequence, localised to a fragmentQuantity
Karl FischerWater content of the solidSolvent content
LAL endotoxinPyrogen load in EU/mgSterility
Sterility testGrowth in defined media over 14 daysEndotoxin, or bioburden count

A mass shift of minus eighteen usually means dehydration or a succinimide intermediate, which is pH-dependent and can be reversible.

Electrospray ionisation soft-ionisation behaviour is well-characterised and standards exist for m/z calibration and mass accuracy assessment.

Always run a blank between samples and check for carry-over.

shareimprove this answerflag
TG
answeredtandem_gradient61k24822 Aug 2025
2Worth adding that the method section is where the answer usually is. – triple_agonist_q 5 months ago
add a comment
Sponsored

Sigma-Aldrich - Certified Reference Materials

Analytical standards and reagents with traceable certificates. Every quantitative result you read inherits the accuracy of the standard behind it.

Shop standards

Your answer

Ask PeptideStack is a static archive. Posting is closed, but the norms are worth stating: answer the question that was asked, show your working, cite the trial or the certificate, and say plainly where the evidence runs out.

Not medical advice. Research-use-only compounds are not approved for human use.