Accepted answer
A 95.2 per cent result against a higher certificate figure is usually a method difference, not a disagreement about the vial. Inter-laboratory spread on identical peptide material runs half a per cent to a full per cent by RP-HPLC. Gradient slope, detection wavelength and integration convention each move the number, and a gentler gradient resolves impurities that a steeper one hides under the main peak — so the better method reports the worse figure. Get the column, the gradient and the wavelength from both VendorInvestigate and the supplier before treating 95.2 per cent as a contradiction.
Purity is a method-dependent figure, and that is not a limitation of the measurement, it is a property of what the measurement actually answers.
Tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.
What each test answers
| Test | Answers | Does NOT answer |
|---|
| RP-HPLC, area % | What fraction of detected material is the target | How much target is present |
| Quantified content | Milligrams of peptide per vial | What the impurities are |
| ESI-MS identity | Whether the molecular weight matches | Purity, or isomeric substitution |
| Peptide mapping | Sequence, localised to a fragment | Quantity |
| Karl Fischer | Water content of the solid | Solvent content |
| LAL endotoxin | Pyrogen load in EU/mg | Sterility |
| Sterility test | Growth in defined media over 14 days | Endotoxin, or bioburden count |
Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.
The Arrhenius relationship for peptide degradation is the basis of accelerated stability testing and also governs how quickly methods drift with temperature.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.
2Adding for future readers: the certificate should carry the lot number, not just a batch code. – marcus_thorbjorn 9 months ago This should be linked from the help pages. – sian_llewellyn 7 months ago add a comment