Accepted answer
It tells you 97.1 per cent of the integrated area came off a wide-pore C18 column where cagrilintide comes off, and the remaining 2.9 per cent did not. That is an area statement at one wavelength, not a mass statement about the vial: 2.9 per cent of area is only 2.9 per cent of mass if every impurity absorbs exactly as strongly as the parent, which none of them do. It also says nothing about how many milligrams are in the glass — water, counter-ion and a short fill are all invisible to it. What a wide-pore C18 column does add is a constraint on what could be hiding: a column that retains by hydrophobicity separates deletion sequences well and separates isomers of identical hydrophobicity not at all.
Put another way, peak shape carries as much information as peak area does, and a badly tailing peak or a shouldered peak is telling you something about the sample or the column that matters.
System suitability checks on replicate injections of a known standard establish whether the method was in control — if the peak area varies by more than two per cent between replicates, something is wrong.
The relevant detail is that acetonitrile is the organic modifier of choice because it has a good UV cutoff, a reasonable viscosity and a refractive index that minimises baseline noise.
Peptide separation by reverse-phase high-performance liquid chromatography is described in the general chapters of the United States Pharmacopeia, European Pharmacopeia and Japanese Pharmacopeia, and the methods converge on essentially the same principles.
The caveat is that HPLC is a purity technique and says almost nothing about whether the main peak is actually your target compound — that is why identity confirmation from mass spectrometry or peptide mapping matters.
The practical summary: specify the method, run the same method on every sample you compare, and use orthogonal techniques to confirm the result.