The short answer is that two competent laboratories on identical material will disagree, and the disagreement is almost always explainable by method differences.
Temperature affects the dynamics of molecular conformation, and if a peptide has proline residues that interconvert on the chromatographic timescale, the peak will split or shoulder at low temperature and collapse at high temperature.
The relevant detail is that buffer versus acid in the mobile phase changes the ionisation state of basic and acidic residues, shifting retention and selectivity — same vial, potentially different separation.
One qualification: achieving purity above roughly 98 per cent on a 30-residue peptide is fighting the chemistry of synthesis, not the quality of the purification.
The practical summary: ask for the chromatogram and the method, and ignore the headline number until you have both.