Concretely, the method matters more than the vial, which is why specifying a method buys you far more than changing suppliers does.
Mobile phase additive choice affects ionisation and peak shape — TFA gives sharp peaks but suppresses mass spectrometry signal, formic acid gives worse peaks but preserves signal.
The relevant detail is that integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.
Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.
Worth noting that method standardisation is poor in the research-grade space compared to pharmaceutical work, so identical-looking methods can produce different results.
If you are ranking vendors, specify a method and have all samples tested at the same place.
7I would gently push back on the second point — the evidence there is thinner than stated. – Dr_Yusuf_Adeyemi 6 months ago 6Adding for future readers: the certificate should carry the lot number, not just a batch code. – bufferline42 5 months ago add a comment