Put another way, read the chromatogram before you read the number, because the number without the trace is not a measurement, it is a claim.
Integration of the shoulder — whether you use perpendicular drop or tangent skim — determines what area gets assigned to the main peak versus the impurity table.
More usefully, tailing factor measures peak shape, and a badly tailing peak spreads into the region where small impurities live, forcing tangent-skim integration that assigns tail area to the main peak.
Proline conformer interconversion kinetics are well-characterised and the half-life is of the same order as the chromatographic peak width at room temperature.
If you are ranking vendors, specify a method and have all samples tested at the same place.
edited 29 Apr 2026 by tobias_maartens — clarified the distinction between purity and content
8Worth flagging that this changed in 2025, so older answers on the site are out of date. – Dr_Nadia_Farsi 5 months ago add a comment