Accepted answer
To be exact about it, the honest answer is that you cannot know for certain what you have without a content assay, and the purity number alone is not enough.
A content assay is always paired with a purity assay because purity tells you what fraction of the measured mass is the target and content tells you the total measured mass.
Reconciling gross mass to label claim
| Component | Typical share | Counted in purity? | Counted in content? |
|---|
| Target peptide | 88–94 % | Yes, as main peak | Yes |
| Related impurities | 1–3 % | Yes, as other peaks | No |
| Counter-ion (TFA or acetate) | 2–8 % | No | No |
| Residual water | 2–6 % | No | No |
| Bulking agent, if present | 0–40 % | No | No |
For peptides at 214 nanometres the response is roughly proportional to the number of peptide bonds, so truncation impurities have lower response factors and overestimate content.
The relative standard deviation on replicate quantitations of a homogeneous sample should be below two per cent when the method is under control.
One qualification: a single result from a single vial is a point estimate, and repeating the assay on a second aliquot is worth doing if the first result is surprising.
If a supplier gives you content without the standard's purity, ask them to provide it.
edited 1 Oct 2025 by tess_amankwah — corrected a unit error in the worked example