More usefully, coupling the chromatography to mass spectrometry adds identity information but trades chromatographic quality for ionisation efficiency.
The 214 nanometre wavelength is chosen because it corresponds to the amide backbone absorption, making response roughly proportional to the number of peptide bonds.
Peptide impurities from solid-phase synthesis include deletion sequences, truncations from premature cleavage, racemised residues from epimerisation and oxidised variants, each of which may have different chromatographic behaviour.
One qualification: the limit of detection on a reversed-phase HPLC is set by the noise on the baseline, which for these molecules is usually in the range of a tenth of one per cent or less, and anything smaller is not reproducibly detectable.
In practice: ask for the chromatogram, check the method section, check the lot number against the vial, and set your accept threshold before you see the result rather than after.
5Adding for future readers: the certificate should carry the lot number, not just a batch code. – Dr_Otto_Lindqvist 7 months ago add a comment