Specifically, peak shape carries as much information as peak area does, and a badly tailing peak or a shouldered peak is telling you something about the sample or the column that matters.
Detector linearity is why a heavily loaded peak can give false purity by compressing the main peak height while leaving impurity shoulders unchanged.
The part that matters: system suitability checks on replicate injections of a known standard establish whether the method was in control — if the peak area varies by more than two per cent between replicates, something is wrong.
The limitation is that co-elution is invisible — if two species happen to have the same retention time, they will report as a single peak at their combined area.
The practical summary: specify the method, run the same method on every sample you compare, and use orthogonal techniques to confirm the result.
edited 20 Aug 2026 by w_okoye — expanded the table to cover the lower concentration
6I have seen exactly this failure mode twice and both times it was the diluent. – bac_or_bust 7 months ago 7The distinction between purity and content cannot be repeated often enough here. – Dr_Colm_Fitzhenry 9 months ago add a comment